Vacuum
Thermopiles and the Measurement of Radiant Energy
C. HAWLEY CARTWRIGHT AND JOHN STRONG
Source: Procedures
in Experimental Physics
by John
Strong
A RADIOMETRIC instrument
consists of a blackened receiver, which is heated by the radiant energy
to be measured. The instrument is provided with some physical means for
measuring the rise in temperature of the receiver produced by the radiant
energy. For the most delicate measurements the means employed must be
responsive to a rise of temperature of the order of a few millionths of
a degree.
In comparison with
other methods of measuring light intensity, a radiometric instrument is
characterized by the direct and simple way in which the response depends
on the intensity of the light; the relation between these two quantities
is linear. Also, the instrument is generally characterized by equal sensitivity
for all wave lengths.
For measuring the
intensity of radiant energy at wave lengths less than 1 ,
radiometric instruments are more reliable but less sensitive than other
instruments such as photoelectric or photographic photometers. Accordingly,
a radiometric instrument is frequently used as a reference instrument
for the calibration of photoelectric and photographic photometers. In
infrared spectroscopy, however, the radiometric instrument is the most
sensitive instrument now available.
When a radiometric
instrument is giving its full response to a beam of light incident on
the receiver, the rate at which the heat is lost by the receiver is in
equilibrium with the rate at which heat is absorbed from the light beam,
. Inasmuch as the heat lost
by the receiver is proportional to the produced rise in temperature, ,
we can write
(1)
where the L's
represent the heat losses in unit time per unit temperature change. Thus,
L4represents the loss of heat by radiation from the
receiver, L2 the loss by air conduction, L3
the loss by conduction through members touching the receiver, and L4
any other means of losing heat, such as, in the case of a thermopile,
Peltier heat loss. Obviously, it is desirable to have the L's small,
and for this reason the energy is to be concentrated onto a small receiver
to reduce L1. Furthermore, the receiver is usually mounted
in a high vacuum in order to make L2 vanish.
The response of the
instrument is determined by the magnitude of ,
and different radiometric instruments are characterized by the manner
in which is measured.
A thermopile measures
by means of one or more thermoelectric
junctions attached to the receiver.1
A microradiometer
measures in the same manner
as a thermopile.2 In this instrument, however, the thermojunctions
and receiver are attached to the moving system of a galvanometer coil,
which is suspended on a fine quartz fiber. The superiority of the microradiometer
over a thermopile lies in the fact that, because no outside lead wires
are required, energy losses in electrical resistance are diminished. However,
the combination of the thermopile and galvanometer makes an instrument
which is awkward to use in a spectrometer, because it must be protected
from vibration in its operating position.
A bolometer consists
of a blackened thin metal strip with electrical connections.3
This strip forms the receiver for the radiations. It is connected as one
arm of a balanced Wheatstone bridge. The change in the electrical resistance
of the strip, as measured by a sensitive bridge galvanometer, is a measure
of .
A radiometer consists
of a system composed of a receiver and a mirror which is mounted in a
partially evacuated case. The system is suspended by a fine quartz fiber.
The back of the receiver is thermally insulated from the front, so that
when a beam of light falls on the receiver, the front is heated more than
the back.4
The radiometer is
most sensitive at a gas pressure of about 0.06 mm of mercury. The gas
molecules which strike the side of the receiver which is warmed by the
radiations leave it with a greater velocity than those which strike the
opposite and cooler side, and therefore a net backward recoil is exerted.
This results in a deflection of the system until the recoil torque is
balanced by the torque arising from torsion of the quartz fiber. The deflection
of the system, as indicated by the mirror, is a measure of the temperature
difference, , between the front
and back surfaces of the receiver.
Anyone interested
in radiometers will find some of the important papers on this subject
listed in the footnotes. One of the features of the radiometer is its
constant sensitivity. This reproducibility of the deflection is due partly
to the use of a quartz suspension but mostly to the fact that the required
pressure (0.06 mm of mercury) is one that is easily maintained permanently
in a closed-off system. The radiometer has been used successfully in the
microphotometer. The application of the radiometer to the microphotometer
places but little demand on flexibility.
When maximum sensitivity
is desired for very delicate measurements, the problem arises of choosing
which type of radiometric instrument will be most sensitive, and further,
which design of a given type will be most sensitive.
There are conflicting
reports on the ultimate sensitivities obtainable with the different types
of radiometric instruments. The thermopile is certainly almost as sensitive
as any other radiometric instrument, and although other instruments might
be made slightly more sensitive than vacuum thermopiles, they are usually
more difficult to construct and use.5 Accordingly, in our treatment
here, the construction details of radiometric instruments other than the
thermopile will be omitted. Vacuum thermopiles are widely used by experimenters
in infrared spectroscopy, possibly more often than all other types of
radiometric instruments taken together.
![](/file/16516/Scientific American - The Amateur Scientist (Tinker's Guild)(2000).iso/amsci01/tblib/thermopiles/thermopiles-01.jpg)
Fig. 1
|
Construction and
evacuation of a sensitive thermopile. The construction of a vacuum
thermopile of the type shown in Fig. 1 will be described here.6
This thermopile has two independent junctions and receivers. Four external
leads are provided, so that these junctions either can be used separately
or can be connected together in series or in opposition. The thermopile
is, made compensating by connecting the junctions in opposition. The receivers
are rectangular and are placed end to end–an arrangement especially
suited to spectroscopic investigations. For special problems the shape
of the receivers as well as other features of the design can, of course,
be altered.
A crystalline quartz
window is attached with Apiezon wax "W." This wax is also used to seal
the other joints Apiezon wax "W" is easy to apply and has an extremely
low vapor pressure–a valuable feature for maintaining a permanent
high vacuum.
![](/file/16516/Scientific American - The Amateur Scientist (Tinker's Guild)(2000).iso/amsci01/tblib/thermopiles/thermopiles-02.jpg)
Fig.
2
|
A porcelain rod of
inch in diameter containing
four holes holds the relatively heavy copper wires on which the thermojunctions
are mounted. The projecting copper wires are fastened together by mica
as shown in Fig. 1 or by Alundum cement as shown in Fig. 2 so that they
will not vibrate. Four flexible and insulated copper leads are soldered
to these heavier copper wires, as shown in Fig. 1, and these are brought
outside the housing through one of the wax seals.
Fig. 3 shows one
method for maintaining a high vacuum of better than 10-4 mm
of mercury in a thermopile. The Pyrex tube shown here is filled with activated
charcoal. The charcoal tube is evacuated and baked for several hours to
outgas it before the stopcock is closed to isolate the system from the
pumps. At first the vacuum will be maintained at better than 10-4
mm of mercury for only a few hours. However, each time the thermopile
is re-evacuated the vacuum lasts longer, so that after about five evacuations,
if the system is tight, the vacuum will remain good for a month or so.
The vacuum is tested by measuring the sensitivity of the thermopile under
some convenient standard condition, such as that of exposing the thermopile
to a 60-watt lamp placed 10 inches away and measuring the response of
the junction with a relatively insensitive galvanometer. The degree of
vacuum obtaining in a thermopile should not be tested with a spark, since
electrostatic forces may destroy the junctions.
![](/file/16516/Scientific American - The Amateur Scientist (Tinker's Guild)(2000).iso/amsci01/tblib/thermopiles/thermopiles-03.jpg)
Fig. 3. (Use
Apiezon wax "W" on the stopcock
|
Wires for the
thermojunctions. One thermoelectric wire is made of pure bismuth,
and the other is an alloy of bismuth and 5 per cent tin. The selection
of this combination of wires to form thermojunctions has been made after
a consideration of the Wiedemann-Franz coefficients, as well as of the
thermoelectric powers of various possible combinations, including such
metals as tellurium and the other bismuth alloys.7
The resistance of
each thermoelectric wire should be at least 10 ohms, and the wire should
not be longer than 3 mm. A bismuth wire 3 mm long with a resistance of
10 ohms has a diameter of about 24 .
The bismuth-tin alloy wire should have about 20 per cent more electrical
resistance than the pure bismuth wire, because of the influence of the
Wiedemann-Franz coefficient. However, owing to the greater specific electrical
resistance of the bismuth alloy wire, its diameter will be about 7
greater than the diameter of the pure bismuth wire.
![](/file/16516/Scientific American - The Amateur Scientist (Tinker's Guild)(2000).iso/amsci01/tblib/thermopiles/thermopiles-04.jpg)
Fig.
4
|
Preparation of
the alloy wires. Thermoelectric wires can be purchased from the Baker
Company, Newark, New Jersey, or they may be prepared by the Taylor process.
To make the wires by the Taylor process, the thermoelectric metal is melted
and sucked up into a thin-wall capillary tube of soft glass. (See Fig.
4.) This tube, containing the metal as a core, is heated in a small electric
furnace and drawn out in the manner shown in Fig. 5. The diameter of the
wires produced in the composite drawn fibers is controlled by the temperature
of the furnace and the speed of drawing. When the temperature of the furnace
is properly regulated, the wires obtained are single crystals which can
be bent and straightened repeatedly without breaking. Wires which are
brittle should be discarded.
![](/file/16516/Scientific American - The Amateur Scientist (Tinker's Guild)(2000).iso/amsci01/tblib/thermopiles/thermopiles-05.jpg)
Fig. 5
|
The glass is removed
from the composite fibers with hydrofluoric acid, which dissolves the
soft glass readily but scarcely corrodes or etches the metal. The hydrofluoric
acid, usually diluted with a little water to suppress fuming, is conveniently
held either in a shallow dish which has been coated with paraffin or simply
in a groove melted in a block of paraffin. The wires are withdrawn from
the acid with metal forceps and washed in a weak solution of Aerosol.8
(Avoid letting the acid come in contact with the fingers.) The wires must
be freed from all glass or difficulty in cutting and soldering will be
encountered. About 5 minutes in the acid is required.
The good wires are
mounted in flat cigar boxes, one for each of the metals. The electrical
resistance of each wire should be measured and its resistance per unit
length noted on a small label attached opposite the wire. After an assortment
of wire sizes has been collected and measured, one is prepared to proceed
with the construction of junctions of prescribed characteristics.
Construction of
the junctions. A microscope of about 10power magnification facilitates
the manipulation and soldering of the thermoelectric wires. An erecting
binocular type giving stereoscopic vision is ideal.
![](/file/16516/Scientific American - The Amateur Scientist (Tinker's Guild)(2000).iso/amsci01/tblib/thermopiles/thermopiles-06.jpg)
Fig.
6
|
Fig. 6 illustrates
the manner of soldering the thermoelectric wires to the copper supporting
wires with a hot tinned sewing needle. The hot-wire device used for heating
the needle is electrically heated, the heat being regulated by a resistance.
The temperature of the tip of the sewing needle can be further controlled
by varying the point of contact between the hot wire and the tip of the
needle.
Wood's metal is used
for soldering. A solution of pure zinc chloride in distilled water is
used as flux. After the soldering is completed, the excess zinc chloride
should be carefully removed with a small brush wet with distilled water.
![](/file/16516/Scientific American - The Amateur Scientist (Tinker's Guild)(2000).iso/amsci01/tblib/thermopiles/thermopiles-07.jpg)
Fig. 7
|
When the thermoelectric
wires, which are selected for size so that each will be about 3 mm long,
are soldered to the tinned copper supports, they are then "cut" to the
proper length by touching them with the hot tinned needle as shown in
Fig. 7. This not only "cuts" the wires but tins their ends at the same
time. Difficulty with this operation will be encountered unless all of
the glass has been These thermoelectric wires are now manipulated with
a cold needle so that their ends are in contact. A little flux is added
to their junction, and the soldering is effected by heat radiated from
the hotwire device. (See Fig. 8.) The junction is to be carefully watched.
The instant to withdraw the heat is indicated by a slight jerk of the
tips of the wires due to surface tension of the fused metal. If the resistance
is too great, each wire is shortened by heating the Wood's metal at the
base of the wire. Molten Wood's metal pulls in the thermocouple wire by
surface tension. The needle is used to heat the Wood's metal. In this
way, it is easy to construct two junctions with only a fraction of an
ohm difference in their
electrical resistances; and, if the wires used have been taken from the
same stock piece of bismuth or alloy wire, the sensitivities of the junction
will match closely.
![](/file/16516/Scientific American - The Amateur Scientist (Tinker's Guild)(2000).iso/amsci01/tblib/thermopiles/thermopiles-08.jpg)
Fig. 8
|
Ruggedness in the
final thermopile is obtained by the use of fine quartz fibers to support
the thermoelectric wires and attached receivers. The quartz fibers are
fastened to the copper supporting wires by thin lacquer as illustrated
in Fig. 9.
![](/file/16516/Scientific American - The Amateur Scientist (Tinker's Guild)(2000).iso/amsci01/tblib/thermopiles/thermopiles-09.jpg)
Fig. 9
|
The receivers are
made of thin gold foil of about 0.5
thickness. This is considerably thicker than sign painters' gold leaf.9
The receivers are cut to size (3 mm by 0.3 mm is a convenient size for
spectroscopy) on the stage of the l0X microscope by means of a razor blade
as shown in Fig. 10. The receivers are strengthened mechanically by giving
them a cylindrical curvature in the following manner: The receiver is
placed on a sheet of thin fine-grade paper mounted on blotting paper,
and a rod of about 0.5 mm in diameter is pressed against it. (See Fig.
10.) Gold is particularly suitable for receivers because it is easily
soldered.
![](/file/16516/Scientific American - The Amateur Scientist (Tinker's Guild)(2000).iso/amsci01/tblib/thermopiles/thermopiles-10.jpg)
Fig.
10
|
A tiny bit of Wood's
metal fused to the junction by radiation and wetted with flux facilitates
attaching the receiver. The gold receiver is laid in contact with the
thermojunction and soldered by heating it with radiation from the hot-wire
device. (See Fig. 11.) A slight jerk of the receiver indicates when the
heater should be withdrawn.
After the receivers
are soldered in place, they are blackened with lampblack or other blackening
material with the aid of a very small amount of glue as a binder. This
mixture is applied to the receiver with a small camel's-hair brush as
shown in Fig. 12.
Finally, two quartz
fibers are fastened over each receiver for added ruggedness. The fibers
are so fine and at the same time such poor heat conductors that the ruggedness
gained by their use more than compensates for the negligible heat leakage
which they introduce. Fig. 13 illustrates the method of securing the receivers
and shows the completed thermopile.
![](/file/16516/Scientific American - The Amateur Scientist (Tinker's Guild)(2000).iso/amsci01/tblib/thermopiles/thermopiles-11.jpg)
Fig. 11
|
Alternative methods
of constructing thermopiles. Some experimenters prefer to make the
housing for a thermopile from blown glass. Fig. 14 shows a popular type
of glass housing. Fig. 15 shows how the junctions are manipulated in the
field of the binocular microscope.
![](/file/16516/Scientific American - The Amateur Scientist (Tinker's Guild)(2000).iso/amsci01/tblib/thermopiles/thermopiles-12.jpg)
Fig.
12
|
The selection of
the proper window material for the thermopile is governed by the spectral
region in which it is to be used. The appropriate choice can be made from
the data given in Table I.
High-melting-point
paraffin for use in the far infrared, listed in Table I, should not be
confused with ordinary low-melting-point paraffin. High-melting-point
paraffin is a crystalline material that does not deform when it is subjected
to small stresses. In order to obtain strength and at the; same time have
the paraffin window very thin, it is advisable to make the window cylindrical.
Fig. 16 illustrates a method of using a tube of paraffin turned out in
the lathe. It is sufficient to have the cylindrical paraffin window only
1 mm thick. Inasmuch as the thermopile cannot be seen through the paraffin
window, it is necessary to adjust the receivers to the focal point of
the radiations with the help of the galvanometer.
|
Although the Taylor
process for preparing thermoelectric wires is recommended, it is possible
to obtain wires by the process used by Professor A. H. Pfund, whereby
the molten metal is splashed on a plate of glass. One may either select
small wires, that are accidentally formed, or cut wires with a razor blade
from the thin foil that is also formed. Wires obtained by this method
have the disadvantage that, owing to fluctuations in their size, it is
difficult to make matched junctions with them.
![](/file/16516/Scientific American - The Amateur Scientist (Tinker's Guild)(2000).iso/amsci01/tblib/thermopiles/thermopiles-13.jpg)
Fig. 13
|
An alternate method
of joining the thermocouple wires and attaching the receivers involves
welding the thermoelectric wires together by means of a condenser discharge.
The details of this procedure are given in the paper cited below.10
The receiver may
be waxed to the welded thermojunction with Apiezon wax "W." This method
of attaching the receiver yields almost the same sensitivity as soldering.
It is easier to construct
a multiple-junction thermopile if one large receiver is waxed to the junctions
than to undertake the delicate task of soldering separate small receivers
to each junction. The electrical insulation between the junctions of a
multiple-junction thermopile can be effected by coating each junction
with lacquer before applying the wax used for holding the receivers.
![](/file/16516/Scientific American - The Amateur Scientist (Tinker's Guild)(2000).iso/amsci01/tblib/thermopiles/thermopiles-14.jpg)
Fig.
14
|
Some experimenters
eonstruct thermocouples in an order almost opposite to that described.
The junction is formed, the receiver is fastened to the junction, and,
finally, the thermocouple is soldered to the supporting wires.ll
This procedure is especially suited to the construction of a thermocouple
with a small circular receiver, such as may be required for stellar radiometry.
For a stellar thermocouple the junction may be soldered with a larger
bit of Wood's metal so that there is formed at the junction a small sphere
of metal, which is then compressed to form a flat receiver of circular
shape and of the desired diameter.
Professor Pfund constructs
thermocouples by compressing the thermoelectric wires together on a plate
of polished steel that is heated to about 100
C.12 The receiver can be joined to the junction in the same
manner. A special device made 5 from a knife switch is used for the manipulation
as shown in Fig. 17.
For most applications
lampblack is suitable for coating the receiver, but in special cases it
may be preferable to use a selective absorbing material for "blackening"
the receiver.l3 Thus, a thermopile used for investigation in
the far infrared spectrum between 52m and 152m , might have receivers
"blackened" with powdered glass. For work in the visible and ultraviolet
spectrum an electrolytic deposit of platinum black is particularly suitable.
![](/file/16516/Scientific American - The Amateur Scientist (Tinker's Guild)(2000).iso/amsci01/tblib/thermopiles/thermopiles-15.jpg)
Fig. 15
|
The loss of heat
by the radiation from a receiver is determined primarily by the emission
of the receiver in the spectral region around l0m (the region in which
the maximum emission from a black body at room temperature occurs). The
emissive power of platinum black in the region around 10m , is weak (about
20 per cent of that of a black body). Thus, the use of platinum black
has the effect of reducing the heat loss L1, so that
the receiver is effectively only one fifth as great as if the receiver
were coated with a material that is black for the heat spectrum as well
as for the visible spectrum. Besides increasing the sensitivity, this
has the further advantage of reducing the theoretical number of junctions
required for the best design. Unblackened silver is suggested for receivers
to be used in the ultraviolet region.
![](/file/16516/Scientific American - The Amateur Scientist (Tinker's Guild)(2000).iso/amsci01/tblib/thermopiles/thermopiles-16.jpg)
Fig.
16
|
Fig. 3 illustrates
the use of active charcoal for maintaining a high vacuum in the thermopile.
An alternate method involves the use of calcium as a getter. This method
has been used by Dr. Pettit of the Mount Wilson Observatory and is quite
satisfactory. Its use amounts to replacing the active charcoal in the
thermopile in Fig. 3 with fresh calcium filings. These calcium filings
are baked out while the tube is connected to the pump. Later, from time
to time when the sensitivity of the thermopile falls off, owing to a decay
of the vacuum, maximum sensitivity can be re-established simply by reheating
the calcium.
![](/file/16516/Scientific American - The Amateur Scientist (Tinker's Guild)(2000).iso/amsci01/tblib/thermopiles/thermopiles-17.jpg)
Fig. 17
|
The use of sensitive
thermopiles. As ordinarily used, the radiant energy focused on the
active receiving surface of the thermopile is interrupted periodically
to isolate the effect of this radiation from the effect of other radiations
falling on the receiver. The excursion of the galvanometer resulting from
interrupting the measured beam is ascribed to changes in the temperature
of the junctions produced by the radiant energy. Considering that delicate
measurements may produce a change in temperature of only 10-6 ░ C.,
it is necessary to interrupt the light rather accurately to compensate
for the first-order drifts which arise owing to a constant warming or
cooling of the surroundings of the entire thermopile. As a result, just
as much time is required for controlling the zero position of the galvanometer
as for determining the deflection produced by the energy being measured.
It is evident that
care is required in selecting the best position for the shutter in an
optical system. For example, it is required that the change in the radiant
energy falling on the thermopile due to closing the shutter should be
the same as the change produced by removing the source of the radiations
without changing the position of any object "seen" by the thermopile.
Otherwise, the variation of radiation from closing the shutter may falsify
the measurement. The shutter is to be put before the entrance slit of
the spectrometer rather than after the exit slit in order to minimize
this possibility.
Compensated thermopiles.
While first-order drifts in the galvanometer can be eliminated even for
an uncompensated thermopile by properly timing the exposures of the thermopile
to the radiant energy, second-order drifts (due to a change in rate of
the drift) can be eliminated only by the use of a compensated thermopile.
In practice, it is difficult to construct a compensating receiver that
will effect the elimination of more than 90 per cent of the galvanometer
drift, but further compensation can be achieved by shunting an electrical
resistance across the most sensitive of the junctions, either the active
or the compensating ones. The junctions to be shunted and the value of
the shunt resistance are determined experimentally. When the shunt resistance
has the proper value, severe temperature changes of the surroundings of
the thermopile housing produce a minimum deflection of the galvanometer.
If care has been taken in constructing a compensated thermopile, the shunting
resistance will be great enough so that the sensitivity of the thermopile
is not appreciably impaired. One method of testing the compensation is
to hold a hot soldering iron a few centimeters in front of the thermopile.
When, for example, a particular thermopile of the type shown in Fig. 1
was compensated, the galvanometer drift was diminished to a twentieth
part of the original drift, and it was reduced further a hundredfold by
the shunting resistance.
Ordinarily, the energy
to be measured is concentrated on one receiver; the compensating receiver
then acts as an external resistance in the galvanometer circuit, and therefore
the deflections are somewhat diminished. In most cases the reduction of
first- and second-order drifts justifies compensation and the attendant
smaller deflections.
By another procedure
the image of the exit slit of the spectrometer covers both receivers,
while a shutter in front of the entrance slit of the spectrometer obscures
first the aperture of the half of the slit focused on one receiver and
then the half focused on the other receiver.l4 Thus the area
of each of the two receivers is half the area of the slit
Theoretically, this
scheme is expected to yield about 40 per cent more sensitivity than the
ordinary compensated thermopile which has the area of the active receiver,
as well as that of the compensating one, each equal to the area of the
slit. In order to realize this 40 per cent gain in another but less desirable
way, the mirror used for concentrating the radiant energy may be tilted
periodically, so that the image of the exit slit of the spectrometer covers
first one receiver and then the other.
Auxiliary apparatus.
Ordinarily a galvanometer having a period of about 7 seconds and a low
resistance of about 10 to 15 ohms is used with a thermopile. For making
delicate measurements, the wires leading from the thermopile to the galvanometer
should be shielded, so that alternating currents will not be induced in
them by stray electromagnetic fields. When the wires are not properly
shielded, induced alternating currents are, in a sense, rectified by the
thermopile, especially by an uncompensated thermopile, and give a spurious
galvanometer deflection.
A simple method of
measuring the galvanometer response is to observe a well-illuminated scale
with a telescope. The galvanometer should be arranged so that the scale
is at a distance of about 5m. A telescope of about 32-power magnification,
placed as close as possible to the galvanometer, should be used. With
a galvanometer mirror 10 mm in diameter, it should be possible to see
the millimeter divisions so clearly on a scale at a distance of 5 m that
deflections on the scale can be estimated to a small fraction of a millimeter.
A lack of definition
is often erroneously attributed to the galvanometer mirror, but it is
usually due to the use of optically imperfect glass for the galvanometer
window. However, there is a limit to the definition attainable, because
of the finite size of the galvanometer mirror and the effect of diffraction.
A simple rule is that the scale distance as measured in meters must not
be greater than the diameter of the galvanometer mirror as measured in
millimeters. Thus, for a scale distance of 5 m, the galvanometer should
be at least 5 mm in diameter. About mm
deflection at a distance of 5 m corresponds to the unavoidable natural
fluctuations in the position of the galvanometer due to Brownian motion.
The accuracy with
which the position of a cross hair on a millimeter scale can be estimated
is much greater than might at first be supposed. A standard laboratory
experiment for students at the University of Berlin is to estimate the
positions of extra marks made on a millimeter scale. All of the extra
marks are made on a ruling engine, so that their positions are accurately
known. Although the lines are all about
mm wide, the student is asked to estimate the position of each extra line
to mm. In estimating these
positions, a student seldom makes an error of
mm, and an experienced observer will have a probable error for a single
reading of about 0.03 mm. Accordingly, it is significant to estimate galvanometer
readings to mm.
![](/file/16516/Scientific American - The Amateur Scientist (Tinker's Guild)(2000).iso/amsci01/tblib/thermopiles/thermopiles-18.jpg)
Fig.
18
|
Fig. 18 shows an
ingenious and accurate arrangement used by Professor Czerny for determining
the magnitude of small galvanometer deflections.15
Relays. A
convenient method of reading galvanometer deflections is to use an optical
amplifier. Also, when it is desirable to record radiometric measurements
photographically, the primary deflections should be amplified by means
of some type of relay, and the deflections of a secondary galvanometer
recorded on moving photographic paper.
The Moll and Burger
thermo-relay may be used for amplifying galvanometer deflections until
Brownian motion becomes conspicuous.l6 Other amplifiers include
the barrier-layer photocell amplifier described by Barnes and Matossi17
and the thermopile with two triangular-shaped receivers described by Cartwright.18
The Barnes and Matossi
type of relay is made by dividing the active surface of a barrier-layer
photocell by scratching along a diameter so as to make two contiguous
semicircular areas of active surface. The arrangement of this amplifier
is illustrated in Fig. 17, Chapter X. Leeds and Northrup produce an amplifying
galvanometer and photocell combination of this type.19
The above methods
of amplifying galvanometer deflections also magnify the drift of the primary
galvanometer. This is undesirable. Pfund and Hardy have devised a resonance
radiometer, which tends to "ignore" drift and separate it from the response
to the measured radiation.20
Their scheme is somewhat
elaborate and requires the use of a tuned pendulum shutter, in addition
to two identical galvanometers. However, the instrument has advantages,
especially when the thermopile is not adequately protected from extraneous
thermal effects. Pfund describes the resonance radiometer briefly as follows:
If primary and secondary
galvanometers are underdamped and adjusted to the same period, then, by
interrupting the radiation falling on the thermopile with a periodicity
corresponding to that of the galvanometers, a condition of resonance is
set up. As a class, resonating systems are characterized by high sensitivity
for "tuned" periodic disturbances and by indifference to random disturbances.
This indifference
to random disturbances unfortunately does not include Brownian motions
of the primary galvanometer. Hardy has measured the effect of the Brownian
motion on the resonance radiometer and has found that the fluctuations
in the deflection of the secondary galvanometer are magnified in accord
with theoretical predictions for fluctuations due to Brownian motion.21
Nevertheless, Hardy feels that delicate measurement to the limit set by
these effects is definitely facilitated by the use of the resonance radiometer.
The slowness of the resonance radiometer (it takes about 90 seconds to
make a measurement) is one of its disadvantages.
Firestone22
has made an ingenious variation from the Pfund scheme. It depends on charging
and discharging a condenser through the secondary galvanometer with a
circuit controlled by the amplified thermocouple current. A photocell
amplifier is used. Naturally, as the output galvanometer circuit has infinite
ohmic resistance, owing to the condenser in the circuit, no net current
can flow, and consequently all deflections are excursions about an unchanging
zero position.
We have emphasized
the importance of using a compensated thermopile to diminish galvanometer
drifts as well as to make the circuit electrically insensitive to high-frequency
electromagnetic radiations. For the most delicate measurements, it is
also necessary to have the galvanometer free from mechanical vibrations.
This can be accomplished by the use of a vibrationless support such as
the type shown in Fig. 19. The description of this vibrationless support
is given in Chapter XIV.
![](/file/16516/Scientific American - The Amateur Scientist (Tinker's Guild)(2000).iso/amsci01/tblib/thermopiles/thermopiles-19.jpg)
Fig. 19. Vibrations
support for a galvanometer. The plywood triangle, on which the device
stands, should be loaded with lead weights until the natural oscillations
have a period of about 2 seconds
|
Construction of
thermojunctions by evaporation and sputtering. There are other applications
of thermopiles and thermocouples, such as their use for vacuum manometers,
for measuring alternating currents, for measuring sound intensities, for
magnifying deflections of a spot of light in thermo-relays, and for total-radiation
pyrometers. We cannot go into all these applications in detail, but the
present chapter and the references cited should serve to guide an experimenter
in these fields. The construction of thermopiles by evaporation and sputtering,
however, warrants a description.
Thermopiles made
from films of the thermoelectrically active metals, produced by evaporation
or sputtering, can be constructed having a very low heat capacity, so
low, in fact, that they will respond to the adiabatic heating produced
by separate sound waves of 5000 cycles frequency.23
One of the metal
films used is bismuth and the other is antimony. The foundation on which
the metal films are deposited must be extremely thin and strong. For this
purpose, glass, mica, or 1acquer films are used.
When a soft-glass
tube is fused at one end and strongly blown out with air pressure so as
to expand and explode a thin bulb, the shattered bulb wall yields fine
ribbons of glass about 1 or 2 mm wide and 1 or 2 cm long. These ribbons
are of such a thickness as to give interference colors and make a suitable
foundation for evaporated thermocouples.
When a mica sheet
is rolled upon a stick of about 2 mm in diameter so that one of the principal
directions is parallel to the stick, it is subject to shearing forces.
These forces produce cleavages, so that when the sheet is subsequently
split, bands from 1 to 0.1 mm wide are obtained which, judging from their
interference colors, are as thin as or thinner than 1m .24
![](/file/16516/Scientific American - The Amateur Scientist (Tinker's Guild)(2000).iso/amsci01/tblib/thermopiles/thermopiles-20.jpg)
Fig.
20.
|
Films for use as
a thermopile base, or for many other purposes, may be made by dropping
a thinned solution of lacquer onto the surface of a bowl of dust-free
distilled water.25 Surface tension causes the drop to spread
out, forming a liquid film on the water over about half the area of the
water surface. The lacquer soon becomes solid as the solvent evaporates.
Fig. 20 shows how these films are taken off the water on a metal frame.
They are allowed to dry after the peripheral area of the film is pulled
back anywhere that it is in contact with the main stretched area. The
thickness of film desired is controlled by varying the dilution of the
lacquer before it is dropped on the water. Extremely thin uniform films
are formed on water cooled to 0░ C. Films as thin as 5 X 10-6
cm are obtainable. Double films formed on a frame as illustrated in Fig.
20 are stronger than single films of double thickness, owing to the fact
that, in the case of double films, weak areas in one film are seldom opposite
weak areas in the second film.
When the thermoelectric
metal is deposited on the foundation film by evaporation, the heat of
condensation of the metal vapor, as well as the heat radiated by the filament
and absorbed by the film, tends to elevate the temperature of the foundation.
It is necessary to prevent the temperature of the film from rising to
a point at which it might be destructive: The films are mounted in the
evaporation chamber in contact with mercury or, better yet, in contact
with a copper cooling block.
Following the procedure
described by Burger and van Cittert26 bismuth and antimony
are used for the thermojunctions, the bismuth being evaporated to form
a strip about 1m thick, while the antimony is evaporated to form a strip
of half this thickness. The proper weight of metal to be evaporated is
determined by a simple calculation using Eq. 2 in Chapter IV. The area
coated with the metal is defined by templates. The bismuth strip, which
is evaporated first, is deposited a little beyond the point which is to
be the center of the junction, say 0.2 mm or so. Then, the evaporated
antimony strip is allowed to overlap the center by an equal amount. The
area where the strips overlap forms the junction. The junction is then
coated by evaporation with bismuth black, antimony black, or zinc black
over a prescribed area, which is defined by baffles.
To form an area to
which electrical contact may be established, gold is sputtered or evaporated
at appropriate points on the metal films. The connector wires may then
be soldered to the gold.
The bismuth crystals
formed in the strip by condensation of vapors have their axes perpendicular
to the base. This crystal orientation results in a thermoelectromotive
force against antimony of 75 microvolts/░ C. The optimum crystal orientation,
so far unattainable by evaporation, gives a thermoelectromotive force
of about twice this value.
Evaporated thermojunctions
are especially useful for making the Moll and Burger type thermo-relay.
Burger and van Cittert were able to obtain a sensitivity about two and
one-half times as great as that obtained with the ordinary rolled Moll
and Burger element.
Considerations
in thermopile design. The thermopile shown in Fig. 1 and described
above can be adapted to meet most of the needs of an experimenter interested
in making radiometric measurements. Some experimenters, especially those
intending to make extremely delicate measurements, will be interested
in the theory for the design of thermopiles. For example, the experimenter
designing a vacuum thermopile of a given area has several decisions to
make. He must decide which metals to select for the thermocouple wires
and determine whether to make few or many junctions. Also, he must decide
on the material to be used for coating the receivers. Or, he may wish
to design a thermopile to operate at atmospheric pressure.
The equations expressing
the theoretical dependence of the galvanometer response on the number
of junctions, area of receiver, characteristics of thermoelectric wires,
and so forth, have been completely developed.27 Calculations
based on this theory require a knowledge of the characteristics of the
thermoelectric wires, namely, their thermoelectric power, electrical conductivity,
and heat conductivity. The calculations also require a knowledge of the
optical properties of receiving surfaces, such as their emissivity and
reflectivity for various wave lengths. With this information, it is possible
to design the thermopile which will give optimum response under the obtaining
conditions.
The characteristic
sensitivity of a thermopile determines its response and, in the theory,
this quantity Q is defined as follows:
(2)
F is the radiant
energy falling on the receivers in unit time, I is the current
in the galvanometer-thermopile circuit, and R is the total resistance
in this circuit. Q is in effect like an efficiency–the efficiency
with which the radiant energy to be measured is converted into galvanometer
deflections.
The expression for
Q for an uncompensated vacuum thermopile of n junctions
in terms of the quantities on which it depends is
(3)
where I is
the thermoelectric current in the thermopile-galvanometer circuit, R
is the total electrical resistance of the circuit, made up of the
thermopile resistance Rt, the galvanometer resistance
Rg, and any external resistance Re.
P is the combined thermoelectric power of the thermoelectric wires,
expressed in volts per degree centigrade. s is the Stefan-Boltzmann radiation
constant, A the area of the receiver, T the absolute temperature
of the receiver, and e its effective radiating power. W1
and W2 are the Wiedemann-Franz coefficients of the two
thermocouple wires.
The quantity in the
brackets represents the total heat losses of the receiver. The middle
term in the brackets represents heat loss by conduction through the wires,
and the third term represents heat loss due to the Peltier effect. Ordinarily
the influence of Peltier heat on the design may be neglected.
The first term in
the brackets represents the heat lost by radiation and gas conduction.
Where the receiver is not in a high vacuum, gas conduction has the same
effect on thermopile design as increasing the magnitude of e and, as we
have pointed out before, the use of a receiver with a small emissivity
for heat radiation, e, has the effect on thermopile design of decreasing
the quantity e A.
Fig. 21 illustrates
for a vacuum thermopile the way in which Q depends on the values
of e A, the number of junctions, and the total electrical resistance
in the thermopile circuit. With e taken as unity the curves are constructed
for A = 1 mm2 and A = 3 mm2. Furthermore,
th'ese curves are for thermoelectric wires made of pure bismuth and wires
of bismuth plus 5 per cent tin having a thermoelectric power of 120 microvolts/░
C. and Wiedemann-Franz coefficients of 3 X 10-8 watt ohm/░
C.2 and 4.2 X 10-8 watt ohm/░ C.2 respectively.
The full curves are for thermopiles having one, two, three, and four junctions,
and the dotted curves are for compensated thermopiles having one and two
~ active and compensating junctions respectively. |
It is desirable,
from a practical point of view, to have a minimum number of junctions
to build. The information given in Fig. 21 facilitates making the compromise
between this practical consideration, on the one hand, and the desire
to have a maximum sensitivity on the other. From curves in this figure,
it is apparent that the energy should be concentrated onto a receiver
which is as small as possible.
By reference to Eq.
3 we see that when the third term in w= the brackets is small in comparison
with the first and second terms, the sensitivity, Q, appears to
be proportional to the thermoelectric power, P. This is not always
the case in practice, and a thermoelectric metal should not be chosen
on the basis of the thermoelectric power alone. As a matter of fact, most
metals with a high thermoelectric power have an unfavorable Wiedemann-Franz
coefficient, which may, in the end, make them even less desirable than
metals such as the bismuth alloys, which are convenient to manage.28
Sensitivity and
rnimmnm energy detectable. When the quantity Q, given by Eq. 2, is
combined with the current sensitivity, dq /dI, and the total
resistance of the circuit, R, it yields the composite sensitivity,
S, of a thermopile and critically damped galvanometer according
to the formula
(4)
Here q is the deflection
of the galvanometer caused by the radiant energy F falling on the receiver
in unit time.
![](/file/16516/Scientific American - The Amateur Scientist (Tinker's Guild)(2000).iso/amsci01/tblib/thermopiles/thermopiles-21.jpg)
Fig. 21
|
It has been customary
to compare the sensitivities, S, of the various radiometric instruments.
This has led to some confusion in the literature. Actually, in making
the most delicate radiometric measurements, we are not interested primarily
in the value of S (which can be made as large as desired by the
use of an amplifier) but rather in the accuracy with which the radiant
energy can be measured in a given time, or, what amounts to the same thing,
in the smallest intensity of radiant energy that can be measured in a
given time with a given accuracy. The magnitude of this smallest deflection
is influenced by disturbances acting on the instruments.
We will designate
this smallest deflection that can be measured by a single reading in a
time t0 and with a mean relative error g, by
the symbol . Until 1926 it
was considered that the elimination of the disturbances on which the value
of depends was simply a matter
of refining experimental technique. Ising was the first to point out that
our experimental technique is already advanced far enough so that in many
cases is determined by the
ever-present Brownian motion fluctuations.29 If we consider
the thermopile system isolated from all disturbances except those produced
by Brownian motion of the galvanometer, then the value of is
easy to determine. According to the principle of the equipartition of
energy, every object with one degree of freedom, such as the moving system
of our galvanometer, will possess a definite amount of kinetic and potential
energy. The average value of the kinetic energy or potential energy at
19░ C. is
watt sec. (5)
The average deflection
due to the potential energy is involved in the expression
Potential energy
= (6)
where K is
the torsional constant of the suspension and k is Boltzmann's constant.
When a reading is taken, the fluctuations of q give rise to an uncertainty
amounting to Therefore, in
order to have a probable error of g, a single deflection must be
at least 1/g times the average fluctuation, or
(7)
It can be shown that
this expression is a general one applicable to any radiometric instrument.
Combining Eq. 7 3 with Eq. 4, we get an expression for the least energy
that can be measured:
watt. (8)
In comparing the
. of different radiometric
instruments, it is necessary to specify not only the accuracy factor g,
but also the time t, to be taken for measuring a deflection. In
the case of a galvanometer, this is because the value of dq /dI
depends on t0. The value of S also depends on
to for other radiometric instruments. It is not correct to assume, as
is usually done, that the value of varies
with the square of the period of the deflecting device. As a matter of
fact, in the case of a thermopile and critically damped galvanometer,
the value of . is proportional
to the square root of the period time of the galvanometer.30
The of
a thermopile and galvanometer can be expressed in terms of the factor
g, the Q of the thermopile, and the period of the galvanometer,
as follows:
watt. (9)
With the values of
Q given by the curves in Fig. 12 it is therefore possible to estimate
the minimum energy falling on the receiver in unit time that can be measured
with a proposed apparatus. It is to be observed that the sensitivity of
the galvanometer does not enter Eq. 9. Eq. 9, however, does imply that
the deflections are measured either directly or with the help of an amplifying
device to the limit set by Brownian motion.
General summary
of the work on thermopile design. The remainder of this chapter will
be devoted to a summary of the results of experimental and theoretical
investigations made by one of the authors, C. Hawley Cartwright, on the
relative merits of the different radiometric instruments, and in addition
will present some general (although not necessarily final) conclusions
resulting from these studies.
Vacuum microradiometers
can be made which will measure less energy, ,
than the best vacuum thermopiles used with a separate galvanometer. This
advantage is not sufficient to offset the practical advantage of greater
flexibility of the thermopile with separate galvanometer.
Vacuum bolometers
and vacuum thermopiles have at present practically the same limit, .
If a material with better characteristics than nickel were available for
the construction of the bolometer strip, this situation would be altered.
Radiometers will
not respond to as small energies, ,
as thermopiles. The direct comparisons made by the author, especially
in Berlin and Brussels, between vacuum thermopiles and radiometers yield
results in favor of vacuum thermopiles. Radiometers are usually much more
sensitive ~ than thermopile and galvanometer combinations, owing to the
use of a much lighter moving system than is possible with a galvanometer.
Brownian motions are, however, increased, so that they more than offset
the advantage of the larger primary deflections.
A question of considerable
importance and one which bears on the above conclusions is the following:
Why is there often considerable variation in the sensitivity of vacuum
thermopiles, in fact, sufficiently large variations to be responsible
for many of the publications that have appeared on improving thermopiles?
The answer is that many vacuum thermopiles are not constructed with the
maximum possible sensitivity, for the following reasons:
1. The sensitivity
of a thermopile depends on the skill exercised in its construction.
2. For the most part,
thermopiles have been constructed without first calculating the proper
design or, if this is done, without dependable information on the physical
properties of the materials used.
3. A sufficiently
high vacuum is not always used. A properly designed and constructed thermopile
should be about twenty times more sensitive in high vacuum than in air,
and, on increasing the vacuum from 10-3 to 10-6
mm of mercury, the sensitivity should be doubled.
4. The thermoelectric
power of the bismuth and bismuthalloy wires is often less than 120 microvolts/░
C. Slight impurities can greatly influence the thermoelectric power of
bismuth by influence on crystal orientation, and so forth. For example,
the thermoelectric power of pure bismuth relative to copper changes from
57 to 107.7 microvolts/░ C. for different crystal orientations.3
5. The influence
of deviations of the properties of bismuth, and especially bismuth alloys,
from the predictions of the Wiedemann-Franz law is generally neglected,
with the result that thermoelectric wires with a resistance which is too
small are used so that the sensitivity falls on the left-hand steep part
of the curves corresponding to those shown in Fig. 21.
Actually, the ultimate
attainable sensitivity for a thermopile is limited by the unfavorable
departure from the Wiedemann-Franz law of the thermoelectric metals that
possess a high thermoelectric power. However, if this were not the case,
it is interesting to note that the thermoelectric power itself would limit
the sensitivity. From Eq. 3 we see that for a thermoelectric power of
250 microvolts/░ C. the heat loss due to the Peltier effect is equal to
the heat loss due to conduction through the wires. Although the possibility
exists of finding better thermoelectric metals than bismuth and the alloy
of bismuth and 5 per cent tin, it seems rather improbable that much progress
will be made in this direction.
It is well to keep
in mind that although tin has ten times less specific electrical resistance
than bismuth, an alloy of bismuth and 5 per cent tin has twice the specific
electrical resistance of pure bismuth. This should be considered when
better thermoelectric metals are being sought. Bismuth itself is an unusually
favorable metal for thermopiles, not only because it has a relatively
high thermoelectric power, but also because it is a pure metal element
having a small specific electrical resistance and does not depart greatly
from the Wiedemann-Franz law.
In order to improve
the sensitivity of thermopiles, there !Y' is the possibility of using
them at low temperatures, where Q can be increased, owing to a greater
thermoelectric power, a more favorable Wiedemann-Franz ratio, and less
radiation loss from the receivers. However, liquid-air thermopiles have
several practical disadvantages.32
1 Brackett,
F. S., and McAlister, E. D., Rev. Sci. Instruments, 1, 191
(1930). Burger,
H. C., and van Cittert, P. H., Zeits. f. Physik, 66, 210 (1930).Coblentz,
W. W., Bureau of Standards, Bull., 11, 131 (1914). Firestone, F.
A., Rev. Sci. Instruments, 1, 630 (1930). Johansen, E. S., Ann.
d. Physik, 33, 517 (1910); Phys. Zeits., 14, 998 (1913). Lebedew,
P., Ann. d. Physik, 9, 209 (1902). Moll, W. J. H., Inaug. Dissertation
Utrecht (1907); Arch. Neerland, 13, 100 (1908). Moll, W. J.
H., and Burger, H. C., Zeits. f. Physik, 32, 575 (1925); Phil.
Mag., 50, 618 to 631 (1925). Paschen, F., Ann. d. Physik, 33,
736 (1910). Pettit, Edison, and Nicholson, Seth B., Astrophys. J.,
56, 327 (1922). Pfund, A. H., Phys. Zeits., 13, 870 (1912).
Rubens, H., Zeits. f. Instrumentenk., 18, 65 (1898).
2 Boys,
C. V., Roy. Soc., Proc., 42, 189 (1887), 44, 96 (1888),
47, 480 (1890); Roy. Soc., Phil. Trans., 180A, 169 (1889).
Coblentz, W. W., Bureau of Standards, Bull., 2, 479 (1906).
Paschen, F., Ann. d; Physik, 48, 272 (1893).
3Langley,
S. P., Am. Acad., Proc., 16, 342 (1881); Annals of the Astrophysical
Obs., 4, 45 (1904), 5, 75 (1905). Leimbach, G., Ann. d.
Physik, 33, 308 (1910).
4Abbott,
C. G., Aitrophys. J., 69, 293 (1929). Coblentz, W. W., Bureau of Standards,
Bull., 4, 391 (1908), 9, 15 (1913). Crookes, Sir William, Roy.
Soc., Phil. Trans., 11, 166, 325 (1876). Sandvik, O., J.O.S.A., 12,
355 (1926). Hettner, G., Zeits. f. Physik, 27, 12 (1924). Nichols,
E. F., Phys. Rev., 4, 297 (1897). Smith, S., Nat. Acad. Sci.,
Proc., 16, 373 (1930). Tear, J. D., Phys. Rev., 23, 641 (1924).
5 Cartwright,
C. H., Physics, 1, 211 (1931). Klumb, Hans, Zeits. f. techn.
Physik, 17, 279 (1936).
6 We wish
to acknowledge the contributions to this design of Professor Firestone
and Mr. Paul Weyrich, of the University of Michigan.
7 Cartwright,
C. H., Zeits. f. Physik, 92, 153 (1934); Ann. d. Physik, 18,
656 (1933).
8 Aerosol
or the detergent Dreft, the latter of which is sold in grocery stores,
has many uses around the laboratory. Besides its usefulness in washing
glass, aluminum mirrors, and so forth, it ean be added to water to decrease
its surface tension and increase wetting power. This is advisable for
washing thermocouple wires, as the solution wets the wires and dissolves
the hydrofluoric acid. Also, for coating the receivers, the solution with
added Dreft has less "attraction" due to surface tension, and accordingly
there is less danger of destroying she work when the brush with its blackening
material is applied.
9 Gold
leaf of the required thickness is prepared by evaporating a proper amount
of gold in vacuum (see "Coating of Surfaces: Evaporation and Sputtering")
from a tungsten eoil onto a glass plate. The film is then washed off the
glass with a stream of water.
10 Cartwright,
C. H., Rev. Sci. Instruments, 3, 73 (1932).
11 Firestone,
F. A., Rev. Sci. Instruments, 1, 630 (1930).
12 Pfund,
A. H., "Radiation Thermopiles," Rev. Sci. Instruments, 8, 417 (1937).
13 Woltersdorf,
W., Zeits. f. Physik, 91, 230 (1934). Forsythe, W. E., Measurement
of Radiant Energy, page 210. New York: McGraw-Hill Book Company, 1937.
Pfund, A. H., J.O.S.A., 23, 375 (1933), 2S, 270 (1933). Strong,
J., Rev. Sci. Intruments, 3, 65 (1932)
14 Badger,
R. M., J.O.S.A., 15, 370 (1927).
15 Czerny,
M., Zeits. f. Physik, 90, 468 (1934). Czerny, M., Heins, H., and
Woltersdorf, W., Zeits. f. Physik, 95, 262 (1935).
16 Moll,
W. J. H., Phil. Mag., 60, 624 (1925). The Moll and Burger thermo-relay
is sold by Kipp and Sonen, Delft, Holland.
17 Barnes,
R. B., and Matossi, R., Zeits. f. Physik, 76, 24 (1932).
18 Cartwright,
C. H., Rev. Sci. Instruments, 5, 221 (1932).
19 Leeds
and Northrup Company, Philadelphia, Pennsylvania.
20 Hardy,
J. D., Rev. Sci. Instruments, 1, 429 (1929), 5, 120 (1934).
Pfund, A. H., Science, 2, 69 (1929).
21 See
also Van Lear. G. A., Jr., Rev. Sc. Instruments, 1, 21 (1938).
22 Firestone,
F. A, Rev. Sci. Instruments, 3, 163 (1932).
23 Harris,
L., and Johnson, E. A., Rev. Sci. Instruments, 5, 153 (1934).
24 This
is the technique described in Burger, H. C., and van Cittert, P. H., Zeits.
f. Physik, 66, 210 (1930).
25 Harris,
L., and Johnson, E. A., Rev. Sci. Instruments, 4, 454 (1933). They
use methyl and ethyl acetate solvent for 2 parts cellulose acetate and
1 part glyptal lacquer at 0░ C. to get the strongest films. Czerny, M.,
and Mollet, P., Zeits. f. Physik, 108, 85 (1937).
26 Burger,
H. C., and van Cittert, P. H., Zeits. f. Physik, 66, 210 (1930).
27 Cartwright,
C. H., Zeits. f. Physik, 92, 153 (1934).
28 Cartwright,
C. H., Ann. d. Physik, 18, 656 (1933). The Wiedemann-Franz coefficient,
W, of any metal can be determined by using the empirical formula
watt ohm/░ C.2,
where is the specific electrical
resistivity and T the absolute temperature. For good conductors
is small, so that W
is the same for all these substances.
29 Ising,
G., Phil. Mag., 1, 827 (1926)
30 Cartwright,
C. H., Physics, 1, 211 (1931). Czerny, M., Ann. d. Physik, 12,
993 (1932).
31 Bridgman,
P. W., Am. Acad., Proc., 63, 347 (1927-1928).
32 Cartwright,
C. H., Rev. Sci. Instruments, 4, 382 (1933).
|